Strobe Alarm: The
white LED to alert the operator when a holiday is
detected.
WARNING:
are susceptible to photosensitive epilepsy or other
photosensitive conditions.
Strobe
On
The strobe alarm can be turned ON or OFF:
Select the strobe icon
state. A red
strobe alarm is turned off.
The strobe alarm setting is retained during power cycles.
Strobe
Off
Reset: Perform a reset of the PosiTest HHD by
selecting
settings and returns the detector to a known condition.
Settings are returned to the following:
Output voltage = 1kV
Audio Alarm = Continuous tone when high voltage is on
and loud alarm when a holiday is detected
Strobe alarm = ON
Power OFF: Select to power-down the PosiTest HHD.
All settings are retained.
Return/Exit Menu: Select to return to a previous menu
or exit the menu.
Wand Mode: When the optional wand accessory is used,
it may be desirable to enable Detection Mode (high
voltage output) without the need to keep the trigger
activated continuously. To activate the Wand Mode press
and hold the power button. Activate the trigger, then
release the trigger while continuing to hold the power
button until a lock icon
the lock icon is displayed, the power button can be
released and the PosiTest HHD will remain in Detection
Mode without the trigger needing to be activated. To
disable Wand Mode, press any button or activate the
trigger.
NOTE:
PosiTest
This visual alert may affect operators who
icon will appear on the display when the
from the Menu. A reset restores factory
11
HHD uses a high-intensity
or
to toggle the strobe alarm
is displayed on the LCD. Once