Transportation And Storage Conditions; Electrical Ratings; Symbols On Product And Rating Plate - KaVo Kerr elements e-motion Instrucciones De Uso

Ocultar thumbs Ver también para elements e-motion:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 333
3 Product description | 3.7 Symbols on product and rating plate

Transportation and storage conditions

Permissible ambient temperature
range
Permissible up to a maximum hu-
midity of
Atmospheric pressure limitation
Mode: intermittent operation
Operating time
Pause time
▶ Do not exceed the threshold load of the motor of 0.5 minutes operating
time / 9 minutes pause.
Note
In practice, pulse loads lasting seconds or pause times lasting seconds or
minutes are realistic, usually without reaching the maximal possible motor
current. This corresponds to the common working procedure of a dental pro-
fessional.
Speed
Speed range of the Motor KL 703
Motor torque
Maximum torque of the
Motor KL 703
Torque of the Motor KL 703 min-
imum

Electrical ratings

Input voltage
Power
Motor cords
Length of connection cord

3.7 Symbols on product and rating plate

The rating plates are affixed to the underside of the unit.
Instructions for use elements e-motion Dental Motor
-20
C to +50
o
5% to 90% non-condensing
700 to 1060 hPa
0.5 minutes
9 minutes
100 - 20,000 rpm (min
2 Ncm
0.1 Ncm
24 V DC
max. 65 W
2.00 m / 78.74"
19 / 55
C / -4
F to +122
F
o
o
o
-1
)

Publicidad

Tabla de contenido
loading

Tabla de contenido