5.2.8 Description of user interface and controls

The TX150/TXF200 features a full colour graphic display, a main dial and two buttons F
and S. All functions (setting temperature, pump speed, countdown timer, presets,
programs settings and standby mode) can be configured from the home screen.
Navigation around the home screen is achieved by rotating the main dial which moves a
white cursor to highlight function icons. Pressing the S button whilst the icon is
highlighted will change the colour of the cursor to red, make the icon active and allow
changes to be made, or in the case of the settings icon, further menus to be displayed.
The primary function of the F button is to exit functions and menus. If F is pressed when
in the home screen whilst the cursor is white then the settings menu is displayed.

5.2.9 Setting the control temperature



.
°c




:
.
Refer to section 4.6 to find the allowable temperatures for each liquid.

5.2.10 Setting a pump speed (TXF200 only)



.
°c




:
.
TX150 & TXF200
Operating Manual
1. Rotate the dial until the set temperature icon is
highlighted, press the S button.
2. Rotate the dial to set the desired temperature.
°c
If no key is pressed for 10 seconds or if F is
pressed, the set temperature icon is no longer
active and will remain at its original value.
3. Press S to store the requested value.
If the temperature selected is higher than the
current liquid temperature the heater will switch
on and the heater icon will be displayed.
The temperature setting range will be limited to
the operating range of the selected liquid.
1. Rotate the dial until the pump speed icon is
highlighted, press the S button.
2. Rotate the dial to set the desired pump speed
°c
over a range of 1 to 5 (1= lowest speed, 5 =
highest speed).
If no key is pressed for 10 seconds or if F is
pressed, the pump speed icon is no longer
active and will remain at its original value.
3. Press S to store the pump speed.
The pump will now operate at the stored pump
speed.
For set temperatures above 150°C, the pump
will automatically operate at a minimum of speed
3.
www.grantinstruments.com