INSTRUCTIONS FOR USE
Intended use: this LED curing light is intended for use by trained dental professionals
Fig.1

ENVIRONMENTAL FACTORS

Storage
Transport
Humidity range
Atm pressure range
INSTALLATION
• Insert light head into the curing light handpiece with a twisting motion. Put the handpiece in the
charging station as shown (Fig.1). Ensure the light head and handpiece are both fitted securely.
• Plug in power transformer connection to the charging station and power supply.
• Station power indicator light will be on if power is connected.
• Once connected, press the ON/OFF switch. Handpiece display screen shows the last selected
time and CH (charging) or FU (charged) in sequence. Handpiece display will automatically turn off
after one minute of charging or not being operated. Battery is still charging even when the display
is off, if the station is connected to power supply. When battery is full, charging will stop automatically.
• The battery is not fully charged prior to shipment. Please charge the battery for a minimum of
4 hours prior to first use.
• To check the battery, press and hold the time switch for 6 seconds, display screen will show:
a.) FU – Battery is full
b.) 4.1 – 3.5 – Battery is normal
c.) Lo – Battery is low
• When battery is low, curing light circuit will automatically compensate for power output or turn the
light off. Always recharge the curing light when the battery is low Lo .
• Plug in transformer to handpiece to recharge directly for easy transportation and mobility if necessary.
10
PRECISION TECH LUMINA
LED curing light
for the purpose of curing dental composites
Light head cover
Heat protective ring
Direct light source head
Protective light shield
On/off switch
Display screen
Time switch
Output switch
Curing light
handpiece
Handpiece charging
station
Power indicator light
SPECIFICATION:
• Radiant intensity: 1200 - 1600 mw/cm²
• Range of wavelength: 440 - 480 nm
• Wavelength peak: 460 mm
• Power supply: Input AC 100-240V, 50/60Hz
Output DC 5V/1A
• Battery: 3.7V, 1100 mAh, Lithium - ion type
• Handpiece dimension: Ø2.7cm x 24cm
• Weight: 160 gms
CONTENTS:
• Curing light handpiece
• Direct light source head
• Disposable curing light sleeves (100 pcs)
• Light head covers (5 pcs)
• Heat protective rings (6 pcs)
• Handpiece charging station
• Protective light shield
• Transformer 100 - 240V
10° C- 40°C
10° C- 40°C
0% - 95%
500 hPa - 1060hPa