6 Maintenance and repair
6.2 Changing the suction motor
20
Changing HEPA filters
If suction power is still weak after replacing the filter bag, then replace the
HEPA filter, too. The HEPA filter is located behind the filter bag.
1. Loosen the two knurled nuts on the maintenance cover.
2. Remove the lid.
3. Remove the filter bag from the nozzle.
4. Then release and remove the two Phillips screws on the perforated
sheet on the inside of the suction system.
5. Take the perforated sheet out and remove the HEPA filter.
6. Put a new HEPA filter on.
7. Put the perforated sheet back on and screw it down with the two
Phillips screws.
8. Put the filter bag back on again.
9. Put the maintenance cover on and screw it down with the two knurled
nuts.
NOTICE
Filter bag fragment
Do not jam the filter bag
Make sure that the cap is sealed properly and the filter bag is not
jammed.
6.2

Changing the suction motor

The motor of the suction device has a service life of approx. 1,200 hrs.
You will notice the motor is worn when suction does not start in either
automatic or manual mode. When this happens, have the motor replaced
by a trained service engineer.
Operating Instructions Suction devices for inLab MC X5 and
Sirona Dental Systems GmbH
64 98 609 D3608
D3608.201.02.02.02 11.2015