CALIBRATION PROCEDURE

The calibration mode should only be accessed by qualified personnel with proper equipment; otherwise calibration integrity
may be compromised. Read the following instructions carefully.
A 2-Point, 3-Point or a 4-point calibration can be performed. A third or fourth point is only necessary when high accuracy is
required at temperatures of 300°F and higher. You must have the proper equipment for every point of calibration.
Do not attempt to calibrate the instrument if the
properly stored to memory.
Refer to API 7.2 or another recognized standard for routine calibration verification recommendations.
Calibration must not be performed in any environment considered to be hazardous.
Equipment needed:
Ice Bath or other low temperature bath with certified reference thermometer.
Warm to hot fluid bath between 20°C (approx 68°F), or higher up to 90°C (approx 194°F) with certified reference
thermometer. (see Note)*
Optional high temperature oil bath at about 150°C/300°F and certified reference thermometer.
*Note for limited calibration: If entire range capability of instrument is not required, the 2 point high adjustment can be made at
a temperature relatively close to the common temperature of the liquid measured and accuracy will be maintained within the
limited range. For example: If liquid product to be measured is commonly less than 38°C (approx. 100°F), then a "high point"
calibration can be made near that temperature. Temperature accuracy above this calibration point cannot be assured.
Calibration Button
11/2012, JK
Figure 1: Probe Assembly Lead Attachment
"Lo
bAtt" has been displayed since the new calibration values may not be

Function Button