The interface line's functions between the laboratory equipment and the PC are selected from
•
the specified signals according to the EIA- RS232C, corresponding with DIN66020 Part 1. The
location of the connector can be seen in Figure 8
Transmission method: Asynchronous signal transmission in start-stop-operation.
•
Mode of transmission: Fully Duplex. 1 start bit; 7 character bits; 1 parity bit [straight (even)); 1
•
stop bit.
Transmission speed: 9600 bit/s
•
When this function is being used, the LCD screen displays "Remote".
•
9. Faults
Instruments cannot be switched ON
•
Check whether the power line is unplugged
Check whether the fuse is broken or loose
Fault during self test
•
Switch OFF the unit, then switch back ON.
Stir speed cannot reach set point
•
Excessive medium viscosity may cause an abnormal motor's speed reduction.
Unit does not turn off when pushing the ON/OFF switch.
•
Check if the residual heat warning function is still ON and the hotplate's temperature is
above 50ºC (the LCD still work and "Hot" flash).
If these faults are not resolved, please contact the manufacturer/supplier.
10. Maintenance and Cleaning
Proper maintenance can keep instruments working properly and lengthen their lifetime.
•
Do not spray cleanser onto the instrument when cleaning.
•
Unplug the power line when cleaning.
•
Only use recommended cleansers:
•
Note:
Forbidden to insert or remove the
RS232C communication line when the
equipment is on!
Dyes
Construction materials Water containing surfactants
Cosmetics
Foodstuffs
Fuels
Isopropyl alcohol
/ Isopropyl alcohol
Water containing surfactants
/ Isopropyl alcohol
Water containing surfactants
Water containing surfactants
Table 4