PicoConnect 442 Differential Probe User's Guide

4. Test procedures

1.
Connect the probe as instructed above.
2.
Run the PicoScope software. Set Channel A to DC coupling and the ±5 V range.
3.
Connect the black test lead and sprung hook to the Cal – output on the back of
the oscilloscope.
4.
Connect the red test lead and sprung hook to the Cal + output on the back of
the oscilloscope.
5.
Verify that the scope display shows a 1 kHz square wave with 4 V peak
amplitude.
5. Specifications
Bandwidth (–3 dB)
DC accuracy
Input
Input resistance
Input capacitance
Attenuation
Differential ranges
Common mode ranges
Measurement category
Connectors
Output
Cable length
Connector
General
Temperature
Humidity
Altitude
Pollution degree
Weight
Enclosure dimensions
Safety approvals
EMC approvals
6
Copyright © 2017 Pico Technology Ltd. All rights reserved.
10 MHz
±2% plus scope accuracy
16.7 MΩ ± 1% each input to scope ground
9.3 pF ± 1 pF each input to scope ground
25:1
±2.5 V, ±5 V, ±12.5 V, ±25 V, ±50 V, ±125 V, ±250 V,
±500 V, ±1000 V (see derating curve below)
125 V (2.5 V to 12.5 V ranges)
1000 V (25 V to 1000 V ranges)
1000 V CAT III (see "Measurement categories" above)
1 x red, 1 x black 4 mm sockets, CAT III shrouded
1.5 m
D9(m) with screw locks. Only for connection to PicoScope
4444 differential oscilloscope.
0 °C to 45 °C operating
15 °C to 30 °C for stated accuracy
–20 °C to 60 °C storage
5% to 80% RH non-condensing, operating
5% to 95% RH non-condensing, storage
Up to 2000 m
2 (as defined in IEC 61010-1:2010: only non-conductive
pollution occurs except that occasionally a temporary
conductivity caused by condensation is expected)
< 200 g
118 x 50 x 28 mm
EN 61010-1:2010
EN 61010-2-030:2010
EN 61326-1:2013
FCC Part 15 Subpart B
(when tested with PicoScope 4444)
DO283-2