Instructions for Use

Installation

Installation

Do not remove the adhesive and protective cap before connecting the turbopump
CAUTION!
to the system.
Figure 3
Install this device indoors only, and in any case do not use the pump
in an environment exposed to atmospheric agents (rain, ice, snow),
dust, aggressive gases, or in explosive environments or those with a
high fire risk.
During operation, the following environmental conditions must be
respected:
In the presence of magnetic fields the pump must be protected using
a ferromagnetic shield. See the appendix "Technical Information" for
detailed information.
The TwisTorr 304 FS pumps must be used in conjunction with one of
the suitable Agilent controller and they must be connected to a
primary pump (see "Technical Information").
232/304
maximum pressure: 2 bar above atmospheric pressure
temperature: from +5 °C to +35 °C
relative humidity: 0 – 90 % (non-condensing)
TwisTorr 304 FS User Manual / 87-901-027-01(B)