Weight Calibration ECF-1
1. Press the
to weighing mode.
2. The display will show CALZ, press the
3. Use ◄, ►, ▲, 0 ~ 9 keys to enter the calibration weight value and then confirm with
the
key.
4. Place the required weight mass onto the scale as indicated on display and press
the
to calibrate the scale.
5. The scale will complete the calibration automatically and will return back to
weighing mode.
Zero Calibration ECF-2
1. Press the
menu and return back to weighing mode.
2. The display will show CALZ, press the
3. The scale will complete the zero calibration automatically and will return back to
weighing mode.
Span Calibration ECF-3
1. Press the
menu and return back to weighing mode.
2. Use ◄, ►, ▲, 0 ~ 9 keys to enter the calibration weight value and then confirm with
the
key.
4. Place the required weight mass onto the scale as indicated on display and press
the
to calibrate the scale.
5. The scale will complete the calibration automatically and will return back to
weighing mode.

FUNCTIONS SETUP

In order to access functions setup, keep pressed the
press the
key.
UF-1 Internal value
1. Press the
2. To continue with other setups use the ► key.
3. To exit menu and return back to weighing mode, press the
key to start ECF-1 mode or the ESC key to exit menu and back
key to start Zero Calibration ECF-2 or the
key to start Span Calibration ECF-3 or the
key to see the internal value.
key.
key.
and without release it,
key to exit the
key to exit the
key.