* When use the test probe to measure, you should place your
fingers at the back of retaining ring.
* Before changing the measuring range, you must ensure that
the test probe is not connected to circuit to be tested.
* For all DC functions, including manual or automatic measuring
range, to avoid the risk of electric shock due to possible
incorrect readings, please use AC function to verify the
existence of any AC voltage. Then, select DC voltage
measuring range equal to or greater than the AC measuring
range.
* Before testing resistance, diode, capacitance measurement
or on-off states, you should first cut off power to the circuit
being tested, and discharge all high voltage capacitors.
* Don't measure resistance or make on-off tests on a live circuit. .
* Before current measurement, you should firstly check the
meter's fuse. Before the meter is connected to the circuit
under testing,you should firstly power off the circuit to be
tested.
* When you make TV repairs or measure power conversion
circuits, you should note the high amplitude voltage pulse
of circuits being tested. The TV filter should be used to
weaken these pulses to avoid the meter damage.
* This meter uses a 9V 6F22 battery. The battery should be
properly installed in the meter's battery compartment.
* When the battery indicator
replaced immediately. Low battery will cause meter reading
errors, and possibly result in electric shock or personal injury
* When you make type III voltage measurement, the voltage
should not exceed 600V; when you make type TV voltage
measurement, the voltage should not exceed 600V.
* When the meter shell (or part of shell) is removed, do not use
the meter.
appears, the battery should be
03
1.1.3 Safety symbols
Symbols used on the meter surface and instructions:
Note-Important safety information, refer to the
instruction manual.
Caution, possibility of electric shock
Equipment protected throughout by double
insulation or reinforced insulation.
Conforms to UL STD. 61010-1, 61010-2-030,
61010-2-033; Certified to CSA STD. C22.2
NO. 61010-1, 61010-2-030, 61010-2-033
Complies with European (EU) safety
standards
Earth (ground) TERMINAL
Direct current
Alternating current
MEASUREMENT CATEGORY III is applicable to test
and measuring circuits connected to the distribution
CAT III
part of the building's low-voltage MAINS installation.
04