Provider'S Notes; Devilbiss Guidance And Manufacturer'S Declaration - DeVilbiss PulseDose PD1000 Guía De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

PROVIDER'S NOTES

No routine calibration or service is required provided the device is used in accordance with the manufacturer's
directions. Between patients wipe with a damp cloth having a maximum 5.25% Sodium Hypochlorite (Bleach)
or 3% Hydrogen peroxide solution. Avoid getting fluids or debris such as sand or dirt inside the oxygen
connections. Do not immerse in water.

DEVILbISS GUIDANCE AND MANUFACTURER'S DECLARATION

WARNING
Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and
put into service according to the Electromagnetic Compatibility [EMC] information provided in the
accompanying documents.
Portable and Mobile RF Communications Equipment can affect Medical Electrical Equipment.
The equipment or system should not be used adjacent to or stacked with other equipment and that if
adjacent or stacked use is necessary, the equipment or system should be observed to verify normal
operation in the configuration in which it will be used.
NOTE– The EMC tables and other guidelines provide information to the customer or user that is essential in
determining the suitability of the Equipment or System for the Electromagnetic Environment of use, and in
managing the Electromagnetic Environment of use to permit the Equipment or System to perform its intended
use without disturbing other Equipment and Systems or non-medical electrical equipment.
Guidance and Manufacturer's Declaration – Electromagnetic Emissions
This device is intended for use in the electromagnetic environment specified below. The customer or the user
of this device should assure that it is used in such an environment.
Emissions Test
RF Emissions CISPR 11 Group 1
RF Emissions CISPR 11 Class B
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations /
flicker emissions
EN - 12
Compliance
Electromagnetic Environment – Guidance
This device uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
This device is suitable for use in all establishments including
N/A
domestic and those directly connected to the public low-
voltage power supply network that supplies buildings used for
domestic purposes.