Trouble shooting
Failure

Possible Cause of Failure

light too stiff
brake pressure of light is set too
tight
light too loose
brake pressure of light is set too
loose
light head lowers
spring pressure in weight
itself
balancer too loose
light flickers
bad contact
light intensity too
wrong voltage supplied
high / low
life cycle of bulb
wrong bulb(s),
too short
wrong voltage supplied
maladjusted
wrong bulb(s)
colours
light area not
wrong bulb(s),
circular
bulb not seated properly,
main lamp broken
glass cover
wrong cleaning agents;
scratched,
glass cleaned or disinfected
frosted or full of
while still hot
cracks
life cycle of
wrong process of sterilisation
sterilisable
handles too short
handles
end of life cycle
damaged
Please contact your responsible Maquet partner, if you encounter the same mistake several times or if
you cannot remove the failure.
56351056 / Index: B
ENGLISH
Suspension / Moveability
Optiical System / Technology
Glass Cover
Sterilisation
Gebrauchsanweisung blue 30/80
Remedy
adjust brake pressure of light head
adjust brake pressure of light head
service
service
have electrician check supply voltage;
check bulb and replace if necessary
check bulb and replace if necessary;
have electrician check supply voltage
replace bulb
check bulb and replace if necessary;
check bulb seat
use suitable cleaning agent;
glass cover to be replaced by service
check sterilisation method
replace handles
Decken- und Wandversion
Ref.
ch. 4.2
ch. 4.2
-
-
ch. 4.1
ch. 4.1
ch. 4.1
ch. 4.1
ch. 3.2
ch. 3.4
ch. 3.4