REPLACING THE BATTERIES

The battery indicator flashes when the batteries are discharged. Replace the battery.
• Using a Phillips screwdriver, remove the battery housing cover.
• Place the 3 batteries (1.5 V "AA") or the 3 storage batteries (1.2 V "AA") in position,
observing the polarities.
• Replace the battery housing cover.
10.

REGULATORY CHECK

10.1. CERTIFICATE OF ADJUSTMENT

New appliances come with a certificate of adjustment. This document indicates the values obtained
during adjustment and certifies that the sensor has been adjusted, in compliance with an in-house
GB
procedure, on a calibration bench with its calibration sensor connected to the International Standard
calibrator.
Tractel
®
recommends an annual metrological check for every appliance.

10.2. ISO 376 CALIBRATION CERTIFICATE

On request, appliances can be supplied with an ISO 376 calibration certificate.
This document certifies, with figures as proof, that the appliance has been calibrated in compliance with
the ISO 376 Standard, on a calibration bench with its calibration sensor connected to the International
Standard calibrator.
This certificate is valid for a maximum period of 26 months.
Tractel
®
recommends an annual metrological check for every appliance.
11.

MAINTENANCE

The LLX1 does not require any special maintenance, except for regular cleaning using a dry cloth.
12.

STORAGE, TRANSPORT, DISPOSAL

Storage: Place the appliance in its original packaging, with the sensor batteries removed. Keep in a
warm, dry place.
Transport: Transport the appliance in its original packaging.
Disposal: Any disposal of the appliance must be carried out in compliance with the regulations in force in
the country of use. For countries subject to European regulations, dynamometers are not covered by
"DEEE" and "RoHS" Directives.