9. Cleaning process
Observe the general safety warnings.
The operator is responsible for the control of the cleaning result.
Take the warming of the cleaning liquid into consideration when cleaning temperature-sensitive items.
Ultrasound can damage sensitive surfaces when operated over a prolonged period of time and particularly when operated at low cleaning
frequencies.
1.
If necessary for the cleaning task at hand, pre-heat the liquid (in units with heating).
2.
When the cleaning liquid has been freshly mixed switch on the Degas / Auto-Degas function for degassing the liquid (approx. 10 min).
3.
Immerse the cleaning items into the cleaning liquid. Do not place any items or containers directly onto the tank floor.
Use either a cleaning basket, a beaker (filled with water + cleaning chemical) or hang the items into the liquid.
4.
Set the cleaning time – switch on the ultrasound.
5.
If required switch on the sweep function (for the intensive cleaning of large items and for robust surfaces).
6.
Monitor the cleaning process: Check the cleaning result by visual inspection. Repeat the cleaning process if necessary.
7.
Rinse the cleaned items after the cleaning and dry if necessary.
10. Maintenance / Care / Repair
Caution! Pull the mains plug before you carry out any maintenance and care works! Do not put the unit under water!
For repair return the unit to your supplier or to the manufacturer.
Electrical safety: Regularly check the housing and the mains cable for damage to avoid electrical accidents.
Care of transducer tank: Lime deposits in the stainless-steel tank can be removed gently by an appropriate cleaning detergent (operate
the unit with water + concentrate). Do not use any scouring media!
Care of housing: Depending on their kind, contaminations can be removed with a mild household cleaner or decalcifier (wipe with wet
cloth).
:
Disinfection
If the unit is operated in the medical and sanitary sector it is necessary to disinfect the transducer tank and the surfaces
regularly for hygienic purposes (standard surface disinfectants).