Running ß-Ketone Control Solution
NOTE: The on-screen Control Symbol flashes on and off repeatedly
until sufficient control solution has been added to the test strip.
(Beeper sounds if enabled.)
5. A ß-ketone quality control test result is available on-screen in
10 seconds. The display does a countdown from 10 to 1.
6. Compare the result on the display with the range printed on
the test strip foil pouch. If the result falls within the range,
your monitor and test strips are working correctly.
7. The result is automatically stored into memory.
8. If test result is above 8 mmol/L for ketone, the screen displays
HI with Ket displayed.