• The Nova Max Test Strips are calibrated against plasma.
• The anticoagulant sodium and lithium heparin may be used.
EDTA is not recommended for use with Nova Max Test Strips.
• Interferences for elevated levels of acetaminophen, tolaza-
mide, uric acid, bilirubin, ephedrine, and methyldopa may
affect results.
• Glucose test results may be falsely low if the patient is severe-
ly dehydrated.
• Critically ill patients should not be tested with the Nova Max
Plus Monitor.
Instructional Notes
1. If in setup mode when the test strip is inserted, the monitor
saves all values entered up to that point and immediately
switches to test mode. Upon exiting test mode the monitor
screen goes blank and does not return to setup mode.
2. If in Data Review mode when the test strip is inserted, the
monitor immediately switches to test mode. Upon exiting
test mode the monitor screen goes blank and does not return
to Data review mode.
3. Battery low icon is displayed in every mode except setup.
4. Once battery level goes below the threshold that triggers the
"low battery" warning, it continues to give the warning until
the monitor becomes unusable due to low battery.