SPECIFICATIONS

POWER
Voltage:
Power:
Fuse:
Line Cord:
Leakage:
MECHANICAL
Height:
Width:
Depth:
Weight:
OPERATIONAL SPECIFICATIONS — FLOW RATES
Range:
Increments:
Accuracy:
Pressure:
Operating Temperature: 50˚F - 104˚F (10˚C- 40˚C)
Operating Humidity:
Operating Altitude:
STORAGE
• Be sure pump dial is in the OFF position.
Storage Temperature:
Storage Humidity:
Storage Altitude:
100 to 130 VAC, 60 Hz, 1-Phase
15 watts, average
200 mA, time-delay type
Hospital-grade, nondetachable (10 feet)
Less than 100 microamps
8.5 inches
6.5 inches
4.8 inches
6.6 pounds
1 - 300 mL/hr
1 mL/hr
±10% or ±0.5 mL/hr, whichever is greater (refer to
CHECKING PUMP ACCURACY, page 12)
Pumps against 18 psi (124 kPa) nominal back pressure
before no FLO alarm.
20% RH - 80% RH, non-condensing
The system is capable of operating as specified at
elevations up to 10,000 feet (3,050 meters) above sea
level.
32˚F - 95˚F (0˚C - 35˚C)
10% RH - 90% RH, non-condensing
The system is able to withstand and operate after exposure
to elevations up to 20,000 feet (6,100 meters) above sea
level.
17