iEi
•
The level switch has responded. There is not enough cooling liquid; top up the cooling
liquid up to approx. 1 cm below the overflow hole (T) in the filling gauge.
Choice of error codes displayed:
•
F1H: sensor broken, sensor resistance over limit
•
F1L: sensor short circuit, sensor resistance below limit
•
F0-: wrong actual value
•
F5-: wrong nominal entry in memory
•
EP-: EEPROM error in memory
If one of these error codes appears, give Heidolph Instruments a call (+49-9122-992068) or
get in touch with your local Heidolph Instruments dealer
ROTACOOL Circulating chiller
Operating temperature range
Resolution/accuracy
Temperature stability
Cooling performance in W at an
ambient temperature of 20°C
Cooling liquid capacity
Refrigerant in the chiller (provided by manufacturer)
Pump's maximum delivery rate
Pump's maximum delivery pressure
Connections for the cooling liquid
Material of the heat exchanger, pump and housing
Basic area (W x D)
Overall height/height of flat surface
Flat surface for rotary evaporator (W x D)
Mains voltage
Power rating
Weight
Protection class
Admissible ambient conditions
WARRANTY / EXCLUSION OF LIABILITY / COPYRIGHT
Warranty
Heidolph Instruments guarantees warranty that the present product shall be free from defects
in material (except wear parts) and workmanship for 3 years from the date shipped off the
manufacturer's warehouse.
Transit damage is excluded from this warranty.
TECHNICAL DATA
[°C]
[K]
[K]
[I]
[l/min]
[bar]
[mm]
[mm]
[mm]
[mm]
[W]
[kg]
18
- 10 to + 40
0,1 / 0,5
+/- 0,5
at 15°C 400 W
at
0°C 300 W
at -10°C 270 W
2,5
R 134 a (CFC free) 0,2 kg
10
0,3
M 16 x 1 / NW 7
stainless steel 1.4301
470 x 580
405 / 95
470 x 405
230V/50Hz or 115V/60Hz
600
35
IP 20
15-32°C bei 80% rel.
Humidity