– If you have just activated the Hold or Tare function, please wait until the scales have
detected a steady weight value. The display will then stop flashing.
– If there has been no activation of a function beforehand, remove the load from the
scales and wait until
... the display
– The ambient temperature of the device is too high or too low. Place the device where
the ambient temperature is between +10 °C and +40 °C. Wait about 15 minutes until
the device has adjusted to the ambient temperature and weigh again.
... the display
and a number appear?
– Use the Start button to switch off the device and start again. After this the device will
function normally.
– If this is not the case, disconnect the power supply by briefly unplugging the AC
adaptor. If this measure is equally unsuccessful, inform the Service department.
...
appears in the height display?
– Slide the head slide up or down quite a ways. This will make the head slide pass a
mark which can be evaluated and the system can now calculate the position correct-
ly again.
8. Servicing
Your seca scale leaves the factory with an accuracy greater than ± 0,15 %. To ensure
that this accuracy is maintained, the product must be carefully set up and regularly main-
tained. Depending on the frequency of use, we recommend servicing every 3 to 5 years.
Please ensure that a qualified service agent is used. If in doubt, contact seca Customer
Service for further information.
Model 763
is displayed, then weigh again.
appears?