Operation; Immunity Tests In Accordance With Iec/En 61000-4-6 - R&S ENY-ITS Manual Del Usuario

Tabla de contenido

Publicidad

®
R&S
ENY-ITS

2 Operation

2.1 Immunity tests in accordance with
IEC/EN 61000-4-6
The standard IEC/EN 61000-4-6 specifies the design and performance of a range of
coupling/decoupling networks (CDNs) for immunity tests in the frequency range from
150 kHz to 80 MHz. Each CDN has a specific design to match the type of cable and
the intended signal carried on the cable.
The impedance stabilization networks (ISN) R&S ENY21 and R&S ENY41 are fulfilling
the requirements of IEC/EN 61000-4-6 for balanced lines hence are suitable for
immunity testing in the frequency range from 150 kHz to 80 MHz. For this purpose,
a 150
150
resistors as well as various adapters for connecting to the ISN are part of the Adapter
Set R&S ENY-ITS.
Fig. 1
Jig CAL U100
Jig TRA U150
SAR T600
SAR T800
SAR T801
Operating Manual 1309.8978.14-01
to 50
adapter (100
termination resistor are required for the calibration of the test system. These
Parts of the R&S ENY-ITS
50 G/150 G Adapter
150 G Termination
Common Mode Adapter for RJ11
Common Mode Adapter for RJ45
Common Mode Adapter for D-Sub 25
series resistor in line with IEC/EN 61000-4-6) and a
Operation
7

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Eny41

Tabla de contenido