2.1.3 Test set-up for calibration

The test generator shall be connected via a 6 dB attenuator to the RF port of the ISN.
The EUT port of the coupling device shall be connected in common mode through the
150 G to 50 G adapter to a power meter with 50 G input impedance. The AE port of the
CDN shall be terminated with 150 G. The set-up for level setting (also called
calibration) is given in Figures 3 and 4.
Fig. 3
Operating Manual 1309.8978.14-01
Equivalent circuit diagram
Test level (e.m.f.)
in V
1
3
10
Test level and measured values
Test set-up for calibration acc. to IEC/EN 61000-4-6 with RJ11 connection
Measured Voltage U
mr
in V
0.1667
0.5
1.667
Operation
Measured Power
in dBm
-2.55
6.99
17.45
9