WARNINGS: Tubing breakage may result in fluid being sprayed from pump. Use appropriate measures to pro-
tect operator and equipment.
Turn off drive before removing or installing Cartridges. Safety guards are provided to minimize risk
of fingers getting caught between the roller mechanism and the base of the module. However, be
safe—Keep your fingers away from these areas.
CAUTION:
Use only tubing/cartridge/pump combinations defined by Tables 1, 2, and 3. Use of other combina-
tions could cause malfunction or damage to the pump.
With the drive stopped, follow this procedure to install Cartridges.
1. Turn the Occlusion Adjustment Knob (see Figure 5) counterclockwise to maximum open. This step is not necessary
when the occlusion is set at the factory setting or if the occlusion has already been set appropriately during previous
pumping.
2. Set the Tubing Retainer according to tubing size (see Figure 7A) when using the large Cartridge. (See Table 4.) No
retainer adjustment is required when using the small Cartridge. (Note: The retainer button located on the Occlusion
Adjustment Knob-side of the large Cartridge is a loose component and may fall out. To reassemble, push it back into
place while pressing in the button.)
3. Select tubing and place in Cartridge (see Figure 7B). If using a Microbore Tubing Set (see Figure 6), place the Collar
up against the Tubing Retainer (see Figure 7C) on both sides of the Cartridge.
FIGURE 5 CARTRIDGE ASSEMBLY
FIGURE 4
REDUCED PULSATION CONNECTION
FIGURE 6 MICROBORE TUBING SET