4. Setting mode

Press both knobs for 3 seconds when heating and stirring are OFF to access the setting mode.
Use the "menu" buttons to browse through the available menus, listed below.
Press the speed control knob to access the selected menu. Then, turn the knob to modify the value of the parameter (when
it is possible). Press the knob to confirm.
To exit the setting mode, do not press any buttons/knobs for 10 seconds or press both knobs at the same time.
Text shown
Display 1
Display 2
TENP
LINt
SPEd
LINt
Strt
NOdE
CtrL
tyPE
PrOb
ALar
PrOb
CAL
tINE
Strt
tINE
SPEd
tINE
HEAt
ANBt
CAL
COun
NOtO
COun
HEAt
UIFI
rSEt
OtC
cFG
StAr
AP
Default value
Range
OFF, 50 ÷ 550 °C
550 °C
10 °C steps
100 ÷ 1700 rpm
1700
100 rpm steps
StOP – run
StOP
FINE
FInE - FASt
On
On - OFF
-10 ÷ +10 °C
0.0 °C
0,1 °C steps
SNAP
SNAP - SEtP
On
On - OFF
OFF
On - OFF
-10 ÷ +10 °C
0.0 °C
0,1 °C steps
---
YES – NO
NO
YES – NO
NO
---
YES – NO
NO
Menu's name and description
Heating plate temperature limit
It limits the maximum value of temperature setpoint for the heating
plate.
If OFF is selected, the heating function is "disabled".
Stirring speed limit
It limits the maximum value of speed setpoint for the motor.
Starting mode
➢ StOP: when the instrument is powered, the heating and the
stirring functions are not active and both displays show OFF
➢ run: when the instrument is powered, it starts working at the last
temperature and speed setpoints
Thermoregulation mode with external probe
➢ FINE: good and optimized temperature control, minimized
overshoot and oscillations, slow rise in temperature
➢ FASt: fast rise in temperature, increased overshoot
External probe safety alarm
It allows to enable / disable AL 6 and AL 7 error messages (see
chap. 5).
Probe calibration
It allows to align the external temperature probe reading to a
reference thermometer.
Starting mode of timer countdown
➢ SNAP: the timer countdown starts as soon as one function
(heating, stirring) is active
➢ SEtP: the timer countdown starts when temperature setpoint is
reached
Keep stirring at timer end
If On is selected, the stirring function remains active when the timer
countdown ends.
Keep heating at timer end
If On is selected, the heating function remains active when the timer
countdown ends.
Heating plate probe calibration
It allows to align the heating plate probe reading to a reference
thermometer.
Motor and heating element operating time
It shows motor and heating element operating times.
Operating times are showed in hours up to 9999 hours (around 416
days). Then they are showed in days (a digits' decimal point is
turned on).
Wi-Fi connection
It allows to enable / disable the Wi-Fi connection.
Reset
It allows to reset all the system settings to the default values.
One time code
It shows the One Time Code during the registration of the device on
the VELP Ermes cloud platform.
Access point
It allows to enable / disable the Access Point mode of the Wi-Fi
connection.