Error And Warning Messages; Velp Ermes Cloud Platform Configuration - Velp Scientifica AREC Connect Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 27

5. Error and warning messages

When the display shows an error message, the instrument's functions are stopped automatically.
AL1
Heating plate overtemperature
AL2
Excessive heating time
The stirring system doesn't run correctly
AL3
AL4
External probe overtemperature
AL6
Slow temperature increase read by the external probe
AL7
Fast temperature decrease read by the external probe
To remove the error message, restart the instrument.
If the alarm persists on the display, please contact VELP Scientifica's technical service department.

6. VELP Ermes cloud platform configuration

VELP Ermes is a revolutionary cloud platform that transforms and improves your laboratory experience.
VELP Ermes creates a connected ecosystem of devices, people and data that transmit information between each other
cutting down distances and expanding your scientific potential.
VELP Ermes collects and stores your data with the maximum level of encryption following the highest cyber-security
standards.
Activate VELP Ermes account:
➢ Go to
www.velp.com
➢ Create your VELP account or login if you already have it
➢ Access to your profile page and select "Configure Your VELP Ermes Account".
Activate your VELP Ermes account after having accepted Terms and Conditions.
➢ You will receive an email containing your ID, Password and a link to proceed to VELP Ermes authentication at
the email address that you entered during the registration
➢ At the first login on VELP Ermes platform, a new password will be requested
Connect the heating magnetic stirrer to your Wi-Fi network:
➢ Turn on the AREC Connect
➢ Enable the Wi-Fi connection and the access point mode as descripted above
➢ Using a PC, tablet or smartphone select the VELP_AREC_(Product Code)_(Serial Number) network available on
the Wi-Fi list
➢ Open a web browser on your PC, tablet or smartphone; insert 192.168.4.1 into the address bar and press "Enter"
Enter "admin" as User ID and "velp_ermes" as Password
➢ Configure the laboratory Wi-Fi network parameters. Then press "Push" to confirm.
NOTE: if necessary, contact your IT specialist to get all your laboratory Wi-Fi connection details
➢ Wait until the Wi-Fi led stops blinking (it has to be constantly ON)
NOTE: if the led is still blinking after 2 minutes, please try to repeat the procedure and check that the network
parameters are correct
Register the heating magnetic stirrer on VELP Ermes platform:
➢ On the AREC Connect, access the One Time Code (OTC) setting menu as descripted above
➢ On the PC, tablet or smartphone, login on VELP Ermes with your account and select "Device Registration"
➢ Enter the AREC Connect product code, serial number and purchase date.
Press "Next"
➢ Enter the One Time Code showed on the display of the AREC.
Press "Next" and wait for the connection
NOTE: if the OTC is not showed on the display of the AREC, please try to repeat the procedure and check that
the network parameters are correct
➢ Give a name to the AREC Connect and enter the address where the instrument is located
Press "Next"
➢ Assign the AREC Connect to an existing virtual laboratory or create a new one through "Laboratory
Management"
At the end of the procedure, you will be able to monitor and control the connected heating magnetic stirrer through
"Instruments" menu.
NOTE: the instrument can be connected only to a 2.4 GHz Wi-Fi network.
NOTE: for further information, please visit VELP website.
through your internet browser
Only if the external probe is connected
Only if the external probe is connected and
if PrOb ALAr is On
5

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

F20500550F20510550

Tabla de contenido