Summary of load bar features
Load bars are interchangeable within a Tru-Test weighing system, with no special configuration changes
•
required.
Standard calibration feature automatically selects pre-programmed load bar calibration.
•
When connected to a Tru-Test indicator, no calibration in the field is necessary. Span calibration can be
•
performed by your Service Centre when applicable.
Any Tru-Test load bars will operate with the latest indicators.
•
Most Tru-Test load bars will operate with the earlier indicators.
•
Calibration
Tru-Test indicators automatically identify and calibrate to the load bars you connect. The Tru-Test indicators
operate with the complete range of Tru-Test load bars. This is known as standard calibration.
There is no requirement to re-calibrate the indicator when swapping from one load bar to another. The standard
calibration feature eliminates the need for time consuming and inconvenient calibration using test weights.
Applications requiring the use of other manufacturers' load bars, which are not pre-programmed, can be satisfied
using span calibration with known test weights (refer to your Tru-Test Service Centre). Once calibrated in this
manner, the indicator can still be used with the Tru-Test load bars using standard calibration. Span calibration
with test weights can also be applied to any Tru-Test load bar if required.