Voltage
dips, short
interruptions
and voltage
variations on
power supply
input lines
IEC61000-4-11
Power
frequency
(50/60Hz)
Magnetic field
IEC61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The SpiroUSB is intended for use in the electromagnetic environment specified below.
The customer or the user of the SpiroUSB should assure that it is used in such an
environment
Immunity
Test
Conducted RF
3 Vrms
IEC61000-4-6
150 kHz to
80 MHz
<5% U
T
(>95 % dip in U
T
For 0.5 cycle
40% U
T
(60 % dip in U
)
T
for 5 cycles
70 % U
T
(30 % dip in U
)
T
for 25 cycles
<5% U
T
(>95 % dip in U
T
For 5 s
3 A/m
IEC 60601
Compliance
test level
Level
Not Applicable
)
(USB
Powered)
)
3 A/m
Electromagnetic Environment -
Guidance
Portable and mobile RF
communications equipment should
be used no closer to any part of the
SpiroUSB, including any cables, than
the recommended separation
distance calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation
distance (d)
3 Vrms
d = 1.2√P
57
Mains power quality
should be that of a typical
commercial or hospital
environment. If the user
of the SpiroUSB requires
continued operation during
power mains interruptions,
it is recommended that the
SpiroUSB be powered from
an uninterruptable power
supply or a battery.
If incorrect operation
occurs, it may be necessary
to position the SpiroUSB
further from sources of
power frequency magnetic
fields or to install magnetic
shielding. The power
frequency magnetic field
should be measured in
the intended installation
location to assure that it
is sufficiently low.