Connecting Controller to Heater and Power Source

1.
When installing the controller, ensure that the controller's thermocouple sensor
makes good contact with both the heater and the object being heated.
a. Consult the heater's installation manual for proper installation of the heater that is
being connected to the controller.
®
2.
Plug heater or optional BriskHeat
output cord into 'Heater Output' receptacle in the
back of the temperature controller.
a. Make sure the plug on the heater or optional output power cord has an AMP Mate-
N-Lock 3- prong plug and is compatible with the AMP Mate-N-Lock receptacle on
the temperature controller.
3.
Connect the receptacle end of the included input power cord into "Power Input" plug
in the back of the temperature controller. Plug the power input cord into a properly
rated power supply.
4.
Flip power switch located in the front to the 'On' position.
Refer to the "Programming Your Controller" section for programming instructions.
Input Power
L2
L1
GND
®
© BriskHeat
Corporation. All rights reserved.
X2 Benchtop
PID Digital Temperature Controller
Heater Output
L1
GND
Thermocouple
L2
RTD
Power Switch
9

PROGRAMMING YOUR CONTROLLER

Basic Functions

Process Temperature (PV
Main
or Setpoint (SP)
Setpoint
SP1
Setpoint
Indicator
Markers

To enter or exit program mode: Press (▲▼) together for 3 seconds

To scroll through functions: Press (▲) or (▼)

▲) together or (
To change levels or options: Press (
*

To view setpoint: Press ( )
*

▲) together
*
To increase setpoint: Press (

▼) together
To decrease setpoint: Press (
*

To reset an alarm or fault condition: Press (▲ ▼) together briefly
When in program mode, after 60 seconds of key inactivity the display will revert to either inPt:nonE or, if the initial config-
uration has been completed, the measured value. Any settings already completed will be retained.
Complete Function List (Levels 1 to 3)
LEVEL 1
Options [Factory
Function
settings] shown
Select Autotune
tunE
[oFF], on, ParK,
Used to switch the Autotune feature on and off, to select ParK or tune at setpoint.
At.Sp
ParK temporarily turns the output(s) off. To use select ParK and exit program mode. To
SP1 Operating Parameters
bAnD
0.1 to * C/°F
SP1 proportional band/Gain or Hysteresis
[10ºC/18ºF]
* 25% sensor maximum Proportional control eliminates the cycling of on-off control.
int.t
oFF, 0.1 to 60
SP1 integral time/reset Auto-corrects proportional control offset error
minutes [5.0]
dEr.t
oFF 1 - 200
SP1 derivate time/rate Suppresses overshoot and speeds response to disturbances
seconds [25]
dAC
0.5 - 5.0 x bAnd
SP1 derivative approach control dAC Tunes warm-up characteristics, independent
[1.5]
of normal operating conditions, by controlling when derivative action starts during
CyC.t
A – –, on.oF,
SP1 proportional cycle-time (see section above)
0.1 - 81 sec
Determines the cycle rate of the output device for proportional control. Select on.oF for
[20]
ON/OFF mode.
oFSt
[0] to * °C/°F
SP1 offset/manual reset
* ±50% bAnd. Applicable in proportional and ON/OFF mode with integral disable:
SP.LK
[oFF] on
Lock main setpoint Locks the setpoint preventing unauthorized adjustment.
®
© BriskHeat
Corporation. All rights reserved.
X2 Benchtop
PID Digital Temperature Controller
Routine Adjustments
View Setpoint
*
*
Increase Setpoint
Second
*
Decrease Setpoint
Setpoint
SP2
Reset Alarm or Fault Message
▲▼
Momentarily Press Together