Adverse Events

Accumulation of biological matter within the
valve can:
Adjusting the valve to a performance setting that
is lower than necessary can lead to excessive CSF
drainage, which can cause subdural hematomas,
slit-like ventricles, and in infants, sunken fontanels.
MRI (Magnetic Resonance Imaging)
Safety Information
The CODMAN CERTAS Tool Kit is considered
"MR Unsafe" in accordance with the American
Society for Testing and Materials (ASTM) Standard
F2503-13.
CAUTION: Do not use the CODMAN CERTAS
Tool Kit in the MR suite.

Detailed Product Description

The CODMAN CERTAS Tool Kit (see Figure 1A)
consists of five components:
These components are packaged within a case
(see Figure 1B) that, when closed, will help
minimize the effect of the CODMAN CERTAS Tool
Kit magnets on other items.
CODFR14061xx02svn.indd 5
CODFR14061xx02svn.indd 5
above the valve. In these instances, verify the
valve setting with x-ray or fluoroscopy. See
SECTION C: Troubleshooting and SECTION D:
Confirming the Current Valve Setting.
cause difficulties adjusting the valve setting
with the Tool Kit
impair the antireflux function
Adjustable Height Locator Tool
Low Profile Locator Tool
Indicator Tool
Adjustment Tool
X-Ray Overlay Tool
5
2015-04-24 16:51:38
2015-04-24 16:51:38