OPERATION
Sweep is frequency modulation of the High Volt current. When you select
They are Continuous (default at 100 pps), 1-10 Hz, 80-150 Hz and 1-150 Hz. The Continuous option allows you to select a continuous
fixed frequency from 1-120 pps.
Frequency is the number of pulses per second of the waveform. To change the Frequency select
the
or
arrows to change the frequency, from 1-120 pps.
up
down
Ramp
Ramp controls the amount of time required to bring the stimulation up to the selected
amplitude. When you select
seconds, 1 second, 2 seconds (default) and 5 seconds.
Display provides two options of viewing output. The options are Voltage (default) and Peak
Current. The ability to assess peak current can help determine tissue response, and an
indication of impedance to current at the electrode skin interface.
23
and press the
sweep
and press
, 4 options are displayed. They are 0.5
ramp
enter
Intelect® Legend Series Combo
button, 4 options are displayed.
enter
and press
frequency
enter
, then use