2 Specification
2.1 Pin allocation of balance output plug D-SUB25P
PIN
Signal
1
1. Limit (LOW)
2
1. Limit (LOW)
3
1. Limit (LOW)
4
2. Limit (OK)
5
2. Limit (OK)
6
2. Limit (OK)
7
3. Limit (HIGH)
8
3. Limit (HIGH)
9
3. Limit (HIGH)
10
4. Limit
11
4. Limit
12
4. Limit
13
5. Limit
14
5. Limit
15
5. Limit
16
ERR
17
ERR
18
ERR
19
During measure-
ment
20
During measure-
ment
21
During measure-
ment
22
Audio signal
23
Audio signal
24
External signal (+)
25
External signal (-)
12
Input/
Output
NO
Output ON, when first limit (LOW) is reached
NC
Output OFF, when first limit (LOW) is reached
COM
-
Joint contact first limit (LOW)
NO
Output ON, when second limit (OK) is reached
NC
Output OFF, when second limit (OK) is reached
COM
-
Joint contact second limit (OK)
NO
Output ON, when third limit (HIGH) is reached
NC
Output OFF, when third limit (HIGH) is reached
COM
-
Joint contact third limit (HIGH)
NO
Output ON, when forth limit is reached
NC
Output OFF, when forth limit is reached
COM
-
Joint contact forth limit
NO
Output ON, when fifth limit is reached
NC
Output OFF, when fifth limit is reached
COM
-
Joint contact fifth limit
NO
Output ON in case of error message
NC
Output OFF in case of error message
COM
-
ERR joint contact
NO
Output ON, if measured quantity is 5 times
greater than reading range
NC
Output OFF, if measured quantity is 5 times
greater than reading range
COM
-
Joint contact during measurement
NO
Output On for audio signal
COM
-
Joint contact when audio signal is given
Input
Contact for output control (+)
Input
Contact for output control (-)
Function
FEJ-A07-IA-defsi-0610