FUNCTIONS

Ambient concentration reset
When entering a contaminated area, it is important to have a function
which allows you to reset the leak detector. Upon initial power on, the
TRITECTOR automatically resets itself to ambient concentration, to ignore
the level of refrigerant present at the tip. Only a level or concentration
greater than the ambient concentration will cause an alarm.

Automatic reset

After 30 seconds, the leak detector resets itself. This function takes into
account the changes in the environment and resets itself just if there are
no significant changes. By default, this function activated, but you have
the possibility to turn this function off. By pressing
&
simultaneously for three seconds this function is disabled.
By disabling this function, the
symbol appears in the display. When
starting the TRITECTOR, it will reset itself for the first time (Ambient
concentration reset).
Manual reset
The manual reset allows you to reset the TRITECTOR whenever you want
and you feel that it is necessary.
To manually reset the TRITECTOR, press briefly
. A double tone
sounds and all rings on the display light up briefly, to indicate that the
manual reset has been performed.
Peak
The Peak function facilitates searching for a leak in a hard to reach area
without being able to see the LCD display.
The Peak function shows the highest concentration detected by the
TRITECTOR. To enable the peak function, press and hold the peak button
for 3 seconds. As long as the function is enabled, the icon
appears in the display.
To disable the function press and hold the peak button again.