Fig. 3-9: RTM Channel1 Probe Menu for RT-ZC10
3.3

Demagnetizing and Zero Adjustment

Risk of circuit damage
● Do not demagnetize while the R&S RT-ZC10 is clamping a conductor to
be measured. Demagnetizing causes current to flow into the conductor,
which may damage parts in the circuit to be measured.
● Check that the conductor being measured is not clamped when supplying
power to the R&S RT-ZC10. When power is turned on, a demagnetizing
waveform is initially applied to the output. This is intentional in the design,
and not a fault.
1. With the waveform measuring instrument input at ground, adjust the trace to the
zero position.
2. Connect the R&S RT-ZC10 current probe and configure the oscilloscope as
described in chapter ​
3. Make sure that there is no conductor in the sensor aperture.
User Manual 1409.7808.02 ─ 01
and chapter ​
3 .1 on page 15
Putting into Operation
Demagnetizing and Zero Adjustment
3 .2 on page
16.
20