R&S
RT-ZC10
● Immediately after powering on, this device may be subject to an appreci-
able offset drift due to the effect of self-heating. To counteract this, allow
the device to warm up for about 30 minutes before carrying out measure-
ments.
● When performing continuous measurements, it is necessary to be aware
that the offset voltage drifts, depending on factors such as the ambient
temperature.
● Under certain circumstances, oscillation may occur while the power supply
is on. This does not indicate a malfunction. Oscillation can be stopped and
operation restored to normal by opening and closing the clamp.
● Acoustic resonance may occur depending on the level and frequency of
the measured current. This does not normally affect measurements unless
a foreign substance such as dust is present on the contact surfaces of the
sensor head.
User Manual 1409.7808.02 ─ 01
Putting into Operation
Connecting the Probe to the DUT
24