Type of error
Battery symbol appears
on the display
Er 0
Er 1
Er 2
Er 3
Er 4
Er 5
Er 6
Er 7
The entire display
is flashing
Er 9
Er A
Er B
The appliance does
not respond when
the keys are pressed
20
TROUBLESHOOTING
Cause
Run-down batteries
The user has moved
during the measurement
Error in the measurement
of the systolic pressure
Error in the measurement
of the diastolic pressure
Inflation blocked because
the wristband is too tight
Air leak or the wristband is
too loose
Inflation above 300 mmHg
Inflation above 15 mmHg
after 3 minutes
Error in accessing the
memory system
Error in the control of
the appliance parameters
Error in the self-diagnosis
of the MCU