For in-the-ear sDemo instruments
General safety information
WARNING
In‑the‑ear sDemo instruments may only be used in
ears with intact skin.
Before every use of an sDemo instrument, ensure
the following using an otoscope and by a visual
inspection:
The ear canal must be free from pathologically
X
modified skin, inflamed skin, or open wounds.
WARNING
Before every use with a client:
Verify the acoustical performance.
X
Visibly check the sDemo instrument for cracks
X
and other damage.
Visibly check the sDemo instrument for
X
persistent contamination that cannot be removed
by cleaning or disinfection.
If you identify any damage, do not use the sDemo
instrument further.
6
WARNING
After every use with a client:
Replace ear piece, receiver wax guard and
X
microphone cover.
Disinfect the sDemo instrument.
X
Instructions for disinfection
In accordance with hygiene regulations, the sDemo
instrument must be cleaned and disinfected after every
use with a client. All parts accessible from the outside
as well as the inside of the battery compartment have
to be treated with an active, sporicidal disinfectant.
Always observe and follow the instructions given by the
manufacturer of the disinfectant.
Remove the battery compartment.
X
Use a soft and clean cloth to remove visual debris
X
from the battery compartment and housing.
Spray disinfectant on all parts accessible from the
X
outside and on the inside of the battery compartment.
Wait at least 5 minutes to allow parts to dry.
X
Install the cleaned battery compartment.
X
If the battery compartment could not be cleaned,
replace it with a new one.
7