S S e e t t s s w w i i t t c c h h i i n n g g o o n n h h y y s s t t e e r r e e s s i i s s ∆ ∆ a a f f o o r r a a l l a a r r m m 1 1
Menu <–> Selection
press the
key to accept the selection
S S e e t t s s w w i i t t c c h h i i n n g g o o f f f f h h y y s s t t e e r r e e s s i i s s ∆ ∆ b b f f o o r r l l i i m m i i t t 1 1
Menu <–> Selection
press the
keey to accept the selection
5 5 . . . . 5 5 . . 1 1 . . 4 4 R R e e s s e e t t L L a a t t c c h h s s i i g g n n a a l l a a t t o o u u t t p p u u t t 1 1
Menu <–> Selection
press the
key
to accept
5 5 . . 5 5 . . 1 1 . . 5 5 S S e e l l e e c c t t S S i i g g n n a a l l F F o o r r m m f f o o r r O O u u t t p p u u t t 1 1
Menu <–> Selection
press the
www.kuebler.com
Example 1.0
Select digit
Set digit
Example: 1,0
Select digit
Set digit, then select signal form for output 1,
Chapter 5.5.1.5
23
Manual reset with red key
Alarm output can only be reset manually if
is selected as the function and is displayed.
Electrical reset with MPI-Input
Alarm output can only be reset electrically, if
is selected as the function and is displayed.
Both manual and electrical reset
Alarm output can either be reset manually via the red key or via a reset pulse
on the MP input.
N N o o t t e e : : The Display Hold function is off.
Positive output signal
Output is a a c c t t i i v v a a t t e e d d , when the instantaneous value > alarm 1
Negative output signal
Output is d d e e a a c c t t i i v v a a t t e e d d , when the instantaneous value > limit 1.
➱
or
,
or
,
Page 23