4.2

Fuse replacement

4.2.1 Mains fuse

To avoid electric shock, disconnect power to the unit and then, pull out the fuser
holder lid on rear panel.
Mains fuse will be 100 mA/250V, FAST, 5 x 200mm.

4.2.2 Input fuses

Refer to Figure 4 and use the following procedure to examine or replace the
meter fuse.
1. Do steps 1 through 2 of the "Battery Replacement" procedure.
2. Using a slot-head scredriver, key, or other appropriate tool.
3. Remove the defective fuse by gently prying one end of the fuse loose and sliding
the fuse out of the fuse holder.
4. Install a new fuse of the same size and rating. Make sure the new fuse is centered
in the fuse holder.
5. Replace the fuse cover in the meter.
The fuses value will be:
F1
-
1 A F 600 V (BUSSMANN BBS-1 or similar)
F2
-
15 A F 600 V (BUSSMANN KTK 15 or similar)
Fuse 2 (15 A 600 V)
Button
Fuse cover
Page 14
Fuse 1 (1 A 600 V)
Figure 4.- Fuse replacement
USER'S MANUAL. MD-200
Opened by screwdriver
Opened by coin
September 2002