6.- ENTERTAIMENT AND MAINTENANCE
6.1.- BACTERIA FILTER CHANGE
The bacterial filter included in your equipment is a closed system filter, It is NON REUSABLE, (GRAVE RISK OF
INFECTION IF USED IN MANY PATIENTS).
6.2.- FUSES CHANGE
If device is connected to the power supply and switch Nº 1 is at position "1", there is not any operation and the switch
is not lighted, check then the fuses condition following up these instructions:
1)
Remove plugs of fuseholders Nº 10 with a screw-driver.
2)
If damaged, replace fuses with news ones (Ref. RC009906)
3)
Replace plugs at original position.
7.- TECHNICAL CHARACTERISTICS
High vacuum aspirator.
Low flow aspirator.
Pump of membrane.
Mean flow: 600 l/h.
Max. vacuum: 80 Kpa (600 mmHg).
Lexan bottle capacity: 1 litre.
Measurements: 340x150x300
Weight: 3,5 Kg. .
Power: 70 VA .
Voltage: 220V / 50 Hz .
Fuses: 2 of 1A, type F.
Equipment of I Class B type.
Average life: 10 years
8.- SPECIAL ACCESSORIES
ON/OFF pedal , with electrical microswitch , and 3,5 meters of electrical cable; Ref. A017078.
Hydrophobic bacteria filter; Ref. FB-193, bag 10 units.
9.- DIAGRAMS AND DRAWINGS
At the request of the customer, ORDISI, S.A. will be supply components lists, description of its, calibration
instructions and whatever necessary information for to repair the equipment by qualified technical personnel. In the
present manual, we herewith enclose the electrical diagram and one complete exploded view of the equipment.
9.1.- PNEUMATIC DIAGRAM
1.- Patient.
2.- Collecting bottle.
3.- Safety valve
4.- Bacteria filter
5.- Vacuum regulator
6.- Vacuum gauge
7.- Motor-pump
8.- Air outlet
NEVER USE YOUR ASPIRATOR WITHOUT BACTERIA FILTER
USING SYMBOLS
Consult accompaniment documents
Protection earth (inside)
Equipment of B class
Year of manufacture
CE registered for notified organization
Selective collection of electrical apparatuse
Página 11 de 13