Explosion Hazard – Revolution Wireless is not designed tor use in areas where an explosion
hazard may occur. An explosion hazard may result from the use of flammable anesthetics,
oxygen-rich environments, skin cleansing agents and disinfectants.
Handle, clean, and dispose of components and accessories that have come in contact with
bodily fluids according to National, Local, and facility rules, regulations and procedures.
Stimulus delivered by the waveforms of this device, in certain configurations, will deliver a charge of 25
microcoulombs (µC) or greater per pulse and may be sufficient to cause electrocution. Electrical current
of this magnitude must not flow through the thorax because it may cause a cardiac arrhythmia.
Users with an implanted neurostimulation device must not be treated with or be in close proximity to
any shortwave diathermy, therapeutic ultrasound diathermy, or laser diathermy anywhere on their body.
Energy from diathermy (shortwave, microwave, ultrasound, and laser) can be transferred through the
implanted neurostimulation system, can cause tissue damage, or can result in severe injury or death.
Injury, damage, or death can occur during diathermy therapy even if the implanted neurostimulation
system is turned off.
The solvents of adhesives and flammable solutions used for cleaning and disinfecting should be allowed
to evaporate before the unit is used.
DO NOT connect the unit to an electrical supply without first verifying that the power supply is the
correct voltage. Incorrect voltage may cause unit damage, malfunction, electrical shock, fire, or personal
injury. Your unit was constructed to operate only at the elctrial voltage specified on the Voltage Rating
and Serial Number plate. Contact your DJO dealer if the unit is not properly rated.
Patient hazard —
- U.S.A federal Law restricts these devices to sale by, or on the order of, a physician or licensed
practitioner. This device should be used only under the continued supervision of a physician or
licensed practitioner.
- Be sure to read all instructions for operation before treating a patient.
- Only authorized individuals are allowed to operate the Revolution Wireless. Individuals are
authorized after receiving training in the operation of the unit and reading this operating on
manual.
- The long-term effects of chronic electrical stimulation are unknown.
- TENS is not effective for pain of central origin (including headache).
- TENS waveforms have no curative value.
i M P o rtA n t i n F o r M At i o n
DANGER!
Warning!