X- Maintenance

• Check the state of the cables regularly.
• If the pump head gets wet or cleaned regularly, lightly lubricate the reels with
oil. Grease the lever.
• Contact Integra MicroFrance if you have any doubts or in the event of
damage.

XI- Changing fuses

Before changing any fuses, turn off power switch and remove the power cord plug
from the electrical outlet.
There are two connection sites for the fuses:
• Site no. 1: Remove fuse holder and change the faulty fuse(s)
(T1AL250VAC)
• Site no. 2: Remove pivoting fuse holder and change the faulty fuse
(TT1.6AL125VAC)
After replacing the fuses, reconnect the device and turn on power switch. If the
device is still not on, check that the fuses are not faulty. If this does not rectify the
situation, please return the device to Integra MicroFrance.

XII- Cleaning and sterilization

• Do not immerse the device or the pneumatic pedal in water.
• Turn off the power and disconnect the device before cleaning.
• Clean the equipment with a phenolic disinfectant or a neutral pH enzymatic
detergent accompanied by water.
• Dry the equipment with a clean, non-abrasive cloth.
• Ensure that the device is completely dry after cleaning. Moisture inside the
device can cause damage.
• Clean the outside of the pneumatic pedal with a phenolic disinfectant.
• Do not try to sterilize the device or the pneumatic pedal.

XIII- Accessories

• The tube is the only accessory which is non-reusable and delivered sterile.
NT010
K – 2015/11