The primary plug is interchangeable. This allows a world wide use. The procedure of changing the plugs is as per
the pictures.
The HEINE
NT300 is intended for installation in medical areas outside the patient environment
®
(at least 1.5 metres from the patient or patient support pursuant to IEC 60601-1, see figure).
To prevent mechanical and electrical breakdown, never install or store the charger in the following areas:
- Areas exposed to high moisture or condensation
- Areas exposed to extreme environmental conditions
- Areas subject to constant vibrations
- Areas subject to severe temperature fluctuations
- Outdoors
Installation
Do not place the NT300 on soft furniture, cloth or carpet.
Do not insert any objects in any opening in the housing.
Fig1: The above figure defines the patient environment (minimum radius around the patient).
Installation, operation and use
The HEINE
NT300 is compatible with the following rechargeable handles:
®
- BETA L rechargeable handle
- BETA NT rechargeable handle
- BETA Slim rechargeable handle
- Standard F.O. laryngoscope handle
- Short F.O. laryngoscope handle
- F.O. angled handle
- Small F.O. laryngoscope handle