cleaning and maintenance
5

Nebulizer Cleaning

All parts of the nebulizer kit, except tubing, should be cleaned according to the following
instructions. Your physician and/or Drive provider may specify a certain cleaning procedure.
If so, follow their recommendations.
To prevent possible risk of infection from contaminated medication,
cleaning of the nebulizer kit is recommended after each aerosol
treatment. Disinfecting is recommended once a day.
Clean After Every Use:
1.
With power switch in the " Off" position, unplug power cord from wall outlet.
2.
Disconnect tubing from the air-inlet connector and set aside.
3.
Disassemble mouthpiece or mask from cup. Open nebulizer by turning cap
counterclockwise and remove baffle.
4.
Wash all items, except tubing, in a hot water/dishwashing detergent solution. Rinse
under hot tap water for 30 seconds to remove detergent residue. Allow to air dry.
5.
For reusable nebulizer kit only, it can be cleaned daily in dishwasher using top shelf
or by following the above cleaning procedure.
The Drive Reusable Nebulizer Kit is dishwasher safe, but do not
place any Drive Disposable Nebulizer parts in an automatic
dishwasher; doing so could cause damage.
Disinfect Daily
1. Using a clean container or bowl, soak items in three parts hot water to one part white
vinegar for 30 minutes OR use a medical bacterial-germicidal disinfectant available through
your provider. Be sure to follow manufacturer's instructions carefully.
2. With clean hands, remove items from disinfectant solution, rinse under hot tap water, and
air dry on a clean paper towel. Store in a zip-lock bag.
NOTE
Do not towel dry nebulizer parts; this could cause contamination.
To prevent possible risk of infection from contaminated cleaning
solutions, always prepare fresh solution for each cleaning cycle
and discard solution after each use.
................................................................................................................
DRIVE COMPRESSOR/NEBULIZER MODEL 18090-BE / 18091-BE
................................................................................................................
WARNING!
CAUTION!
WARNING!
11