ELECTROMAGNETIC COMPATIBILITY
Guidance and manufacture's declaration - electromagnetic emissions
This device is intended for use in the electromagnetic environment specified below. The customer or the user of this device should assure that it is used in such an environment.
Emmission Test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/flicker emissions
IEC 61000-3-3
Guidance and manufacture's declaration - electromagnetic immunity
This device is intended for use in the electromagnetic environment specified below. The customer or the user of this device should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level
Electrostatic discharge (ESD)
± 6 kV contact
IEC 61000-4-2
± 8 kV air
Electrical fast
transient/burst
± 2 kV for power supply lines
IEC 61000-4-4
± 1 kV for input/output lines
Surge
IEC 61000-4-5
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
Voltage dips, short
<5 % UT
interruptions and voltage
(>95 % dip in UT)
variations
for 0.5 cycle
on power supply
40 % UT
input lines
(60 % dip in UT)
IEC 61000-4-11
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 s
Power frequency (50/60 Hz)
3 A/m
magnetic field
IEC 61000-4-8
18 • ENGLISH
For all ME EQUIPMENT and ME SYSTEMS
Compliance
Electromagnetic Environment Guidance
Group 1
This device uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
This device is suitable for use in all establishments, including domestic establishments and those directly connected to
Class B
the public low-voltage power supply network that supplies buildings used for domestic purposes.
Class A
Complies
For all ME EQUIPMENT and ME SYSTEMS
Compliance Level
± 6 kV contact
± 8 kV air
± 2 kV for power supply lines
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
<5 % UT
(>95 % dip in UT)
for 0.5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 s
3 A/m
Table 1
Table 2
Electromagnetic Environment Guidance
Floors should be wood, concrete or ceramic tile. If floors are covered with
synthetic material, the relative humidity should be at least 30 %.
Mains power quality should be that of a typical commercial or hospital
environment.
Mains power quality should be that of a typical commercial or hospital
environment.
Mains power quality should be that of a typical commercial or hospital
environment. If the user of this device requires continued operation during
power mains interruptions, it is recommended that this device be powered
from an uninterruptible power supply or a battery.
Power frequency magnetic fields should be at levels characteristic of a
typical location in a typical commercial or hospital environment.