2500KL ASSEMBLY INSTRUCTIONS (CONTINUED)
Assembling the 2500KL Scale (Refer to Figures 1-7)
Caution: It is recommended that the assembly be performed by two persons.
1. Place scale platform on a flat, level, and dry surface.
2. Locate and pick up the handrail. Find the load cell cable, located at the bottom of the
handrail.
3. Untie the load cell cable, and carefully pull the load cell cable end out the handrail bottom.
4. Position the handrail over the left and right handrail mounting holes array on the platform
(see Figure 3).
5. Feed the cable through the large center hole of the handrail mounting holes (located on
the right side of the "Health o meter" logo). Be sure not to pinch the cable.
6. Fasten the handrail to the platform using the (8) M8 1.25 x ¾ screws (A), (8) M8 washers
(D) and the included 5mm Allen wrench (see Figure 1).
Caution: Do not over-tighten the screws.
7. Pull forward on the handrail to tip the scale over, and gently place it face down on the
floor.
Caution: Take care to hold the ramp down when moving the scale - the ramp can flip
open when the scale is tipped
8. Grasp the load cell cable protruding from the bottom of the platform and feed it through
the protective metal shield. Plug the cable's load cell connector into the load cell port (see
Figure 4).
9. Pull the excess cable through the top of the handrail.
10. Thread the (4) feet into the underside of the platform. Each of the feet should be
completely hand tightened. Back feet out 1-½ full turns. This will allow for proper load cell
operation and scale accuracy (see Figure 5).
11. Carefully tilt the scale back to an upright position on the floor.
Caution: Take care to hold the ramp down when moving the scale - the ramp can flip
open when the scale is tipped
Warning: Do not drop the scale on its feet. This can result in permanent load cell
damage.
12. Adjust the feet to level the scale. Be certain that the feet are not completely tightened
into the base assembly (refer to the scale platform - feet assembly instruction above).
13. Rotate the display support so that it is perpendicular to the handrail (see Figure 6).
Position the display assembly close to the display support. Plug the load cell cable
connector into the load cell port on the rear of the display assembly. The port is labeled
"LOAD CELL" (see Figure 7).
Note: Do not attach the AC adapter power cord into the back of the display assembly at this
point.
forward.
back.
7