Tube Replacement - Kronos pH/ORP Serie Manual De Instalación

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

Priming
With the pump in Stop (press the
speed for 30 seconds. To stop the pump before the 30 seconds passed, press again the

Tube replacement

With the pump in Stop (press the
to operate at minimum speed for 60 seconds, with operation and pause intervals, to allow the tube replacement. To
stop the pump before the 60 seconds passed, press again the
Alarms
Dislay
Alarm LED on
"Lev" icon flashing
Alarm LED on
"Lev" and "Stop" icons flashing
Alarm LED on
"Flw" icon flashing
"OFA" icon flashing
"Stop" icon flashing
"Alm" icon flashing
"Cal" icon flashing
Flashing words Mtr and Stop
Flashing words Brk and Stop
0000138014
key), by pressing the
key), by holding down for at least 3 seconds the
Cause
End level alarm, without pump
operation interruption.
End
level
alarm,
operation interruption.
Flow alarm activated, the pump has
not received from the flow sensor
the programmed number of signals.
O.F.A. alarm
The value read by the probe is out
of the alarm band set
Probe not calibrated alarm
Motor block alarm.
Tube break alarm.
rev. 1.0 EN
key, the pump will start to operate at maximum
key.
Restore the liquid level.
with
pump
Restore the liquid level.
Press the
Press the
flashing of the "Stop" icon, press
the key again to restart the pump.
Check the "Alarm Band" parameter
for
the
programming mode.
Perform
procedure.
Power cycle the system, if the alarm
persists check the installation or the tube
position.
Reset the tube.
key.
key, the pump will start
Remedy
key.
key to stop the
correct
settings
the
probe
calibration
in
11