Put On Tray; Set Up And Align Scale - Seca 745 Manual De Instrucciones

Ocultar thumbs Ver también para 745:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 89

5.3 Put on tray

1.
3.
2.

5.4 Set up and align scale

Spirit level
Air bubble
34 •
1. Place the tray centrally on the scale.
2. Ensure that the locking levers for the tray engage in
the openings in the housing.
3. Push the locking levers down.
The tray is attached to the housing.
NOTICE!
Incorrect measurement due to force shunts
If the scale is set up or aligned incorrectly,
weight will not be measured correctly.
– Place the scale on a firm and level surface.
– Do not place the scale on changing mats or
towels.
– Set up the scale so that it is not in contact
with any walls or objects.
1. Place the scale on a firm and level surface.
2. Verified models: align the scale by turning its feet so
that the air bubble is in the exact centre of the spirit
level.
NOTE
Alignment of the scale must be checked and
corrected every time the location of the scale is
changed.