Message
54
What it means
The sample was
applied before the
meter was ready.
One of the following
may apply:
You may have high
glucose and have
tested in an
environment near
the low end of the
system's operating
temperature range
(50–111°F).
or,
What to do
Repeat the test with a new test
strip. Apply a blood or control
solution sample only after
or
APPLY SAMPLE
APPLY CONTROL
appears on the display. If this
message continues to appear,
contact Customer Service.
If you tested in a cool
environment, repeat the test in a
warmer environment with a new
test strip; see pages 24–27.
If the error message appears
again, contact Customer Service.