Requirements and sterilization cycle according to the
equipment:
- Equipment: autoclave
- Wet heat sterilization: 134ºC at 2 bar (200 KPa) for
18 minutes
Important: metal parts of different types should not
touch one another during sterilization. Any such
contact would produce electrolytic couples that
would cause local damage. To avoid this
phenomenon, place the objects in a cloth or a micro-
porous sterilizing bag.
5. 3 FUSE REPLACEMENT
Proceed as follows:
1- Turn the ON/OFF button to "O" (23) and unplug
the mains lead from its socket (24).
2- Insert the tip of a screwdriver into the notch (25)
to open the protective cap.
3- Replace the fuses (26) with fuses of the same
rating and type.
4- Push the fuse compartment back into its housing.
The S
®
designed by SATELEC is a medical
ERVOTOME
device of class IIb (European Directive no. 93/42/EEC
concerning medical devices; Transposition to the
French national legislation by Decree no. 95/292 of
16/03/1995).By application of the French Decree of 5
December 2001 and the corresponding Order of 3
March 2003 concerning the obligation to ensure
proper maintenance and quality control of medical
devices, the operator -who is responsible for due
implementation of the applicable maintenance
operations, is invited to refer to, and to apply the
maintenance operations as introduced in the
following standards:
CEI 1289-1: High-frequency surgical equipment - Part
1: Operation
CEI 1289-2: High-frequency surgical equipment - Part
2: Maintenance.
VI - TROUBLESHOOTING
Please refer to the table on the following page