Operation
Forehead Temperature Analysis
1. Before measuring forehead temperature, make sure
there is no water, sweat, or direct sunlight on the
forehead.
2. Press and hold 'F1' while slowly scanning the entire
forehead from one temple to the other. Make sure to
keep the thermometer in contact with the forehead
at all times while moving from one side of the
forehead to another.
3. Release the 'F1' button. The temperature results will
appear on the LCD display. A second measurement
may be taken once the LCD backlight turns off.
NOTE: K eeping the infrared scanner at one point may
affect the accuracy of temperature results. Make sure
to move the thermometer across and in contact with
the forehead for better temperature result accuracy.
Eardrum Temperature Analysis
1. Remove the sensor plastic cover.
2. Before measuring eardrum temperature, clean out
the ear canal.
3. Gently insert the sensor tip fully into the ear canal.
4. Press 'F2' while holding the sensor in the ear canal.
The thermometer will chime and the temperature
results will appear on the LCD display.
9
EN